Pressure Sensor Bridge Output
Calculate bridge output voltage for piezoresistive pressure sensors. Determine mV output from excitation voltage, sensitivity, and applied pressure.
Linear ratiometric bridge · V = V_ex·S·P/P_FSAPI slug pressure-bridge-output
Sensor and circuit
Convert
0 kPa to 100 000 kPa
- Bridge Output
- 25mV
- Amplifier Output
- 1.5V
- ADC Code
- 1 861
- 0x745
- Full-Scale Output
- 50mV
- Fractional Deflection
- 50.0%
- Resolution
- 0.02686kPa/LSB
Transfer curves
Resolution 0.02686 kPa/LSB across the rangeLookup table
Every 5 kPa from 0 kPa to 100 kPa| Applied Pressure | Bridge Output | Amplifier Output | ADC Code |
|---|---|---|---|
| 0kPa | 0mV | 0V | 0 |
| 5kPa | 2.5mV | 150mV | 186 |
| 10kPa | 5mV | 300mV | 372 |
| 15kPa | 7.5mV | 450mV | 558 |
| 20kPa | 10mV | 600mV | 744 |
| 25kPa | 12.5mV | 750mV | 930 |
| 30kPa | 15mV | 900mV | 1 117 |
| 35kPa | 17.5mV | 1.05V | 1 303 |
| 40kPa | 20mV | 1.2V | 1 489 |
| 45kPa | 22.5mV | 1.35V | 1 675 |
| 50kPa | 25mV | 1.5V | 1 861 |
| 55kPa | 27.5mV | 1.65V | 2 048 |
| 60kPa | 30mV | 1.8V | 2 234 |
| 65kPa | 32.5mV | 1.95V | 2 420 |
| 70kPa | 35mV | 2.1V | 2 606 |
| 75kPa | 37.5mV | 2.25V | 2 792 |
| 80kPa | 40mV | 2.4V | 2 978 |
| 85kPa | 42.5mV | 2.55V | 3 165 |
| 90kPa | 45mV | 2.7V | 3 351 |
| 95kPa | 47.5mV | 2.85V | 3 537 |
| 100kPa | 50mV | 3V | 3 723 |
Formula
| Symbol | Quantity | Your value |
|---|---|---|
| S | Sensitivity (mV/V) | 10 mV/V |
| PFS | Full-scale pressure (kPa) | 100 kPa |
Assumptions
6 items
- 01The sensor is linear and ratiometric, V_out = V_ex × S × P/P_FS, with S the full-scale output per volt of excitation; zero offset, span tolerance, non-linearity, hysteresis and temperature effects are not modelled.
- 02Pressure is in the unit of the full-scale pressure, kPa on the page, and is not limited to full scale, so a pressure above P_FS reads more than 100 % deflection.
- 03The excitation is an ideal source, and the amplifier an ideal gain G with no offset, noise or bandwidth limit that does not load the bridge.
- 04The ADC is an ideal unipolar N-bit converter with full scale V_ref and no INL, DNL or noise: code = floor(V_amp/V_ref × 2^N), held between 0 and 2^N − 1, so an output below zero reads 0 and one at or above V_ref reads 2^N − 1.
- 05A code typed in is read at the middle of its step, (code + 0.5)/2^N × V_ref, and solved back through the same model.
- 06Resolution is V_ref × P_FS/(2^N × G × V_ex × S × 10⁻³) per step with S in mV/V, the same at every pressure because the model is linear.
Worked example
100 kPa sensor at 60 kPa
An MPX2100-class 100 kPa piezoresistive sensor gives 4 mV/V at full scale. Excited at 5 V and read through a gain-150 amplifier into a 3.3 V, 12-bit ADC, it sees 60 kPa.
| Step | Expression | Result |
|---|---|---|
| Full-scale output | 4 mV/V × 5 V | 20 mV |
| Fraction of full scale | 60 kPa / 100 kPa | 60.0 % |
| Bridge output | 20 mV × 60.0 % | 12 mV |
| Amplifier output | 150 × 12 mV | 1.8 V |
| ADC code | floor(1.8 / 3.3 × 2^12) | 2 234 |
| Resolution | 3.3 V × 100 kPa / (2^12 × 150 × 20 mV) | 0.02686 kPa/LSB |
How it works
This calculator computes piezoresistive pressure sensor bridge output voltage, essential for process control engineers, HVAC system designers, and automotive sensor integrators. Piezoresistive pressure sensors contain a Wheatstone bridge of diffused or thin-film strain gauges on a silicon or steel diaphragm. Applied pressure deflects the diaphragm, changing resistance to unbalance the bridge. Output is Vout = Vex * S * (P/Pfs), where Vex is excitation voltage (5-10V typical), S is sensitivity in mV/V (typically 10-30 mV/V per OIML recommendations), P is applied pressure, and Pfs is full-scale pressure. A 20 mV/V sensor on 5V excitation produces 100 mV at full scale. Per IEC 61298, pressure sensor specifications include accuracy (+/-0.1 to +/-1% FS), nonlinearity (+/-0.1-0.5% FS), hysteresis (+/-0.05-0.2% FS), and thermal zero shift (typically +/-0.02% FS/C). Industrial sensors from Honeywell, Sensata, and Bosch achieve total error band (TEB) of +/-0.25% FS over -40 to +125C per AEC-Q100 automotive qualification.
Practical tips
- Use ratiometric operation: connect both ADC reference and sensor excitation to the same regulated voltage; if supply fluctuates +/-5%, both scale proportionally and ratio Vout/Vex remains constant per Honeywell Technical Note HSC-AN-800
- For absolute accuracy, perform two-point calibration at zero pressure and known reference pressure (NIST-traceable calibrator) to correct for both offset and gain errors per ISO 17025 requirements
- Add 100 nF ceramic capacitors from each excitation line to ground, close to the sensor, to filter high-frequency noise from PWM switching that would appear as measurement noise
Common mistakes
- Applying excitation exceeding sensor maximum: overvoltage causes bridge self-heating, shifting zero by 0.1-1% FS and degrading accuracy; verify maximum excitation (typically 5-12V) per manufacturer datasheet
- Installing sensor upside-down relative to calibrated orientation: many sensors include diaphragm dead-weight in zero calibration; orientation change causes offset shift equal to diaphragm pressure head (0.1-1% FS for liquid-filled sensors)
- Neglecting common-mode voltage at amplifier input: bridge output rides on Vex/2 common mode; choose INA with input range including Vex/2 on your supply rails per Texas Instruments INA128 datasheet
Frequently asked questions
3 questions
What is the difference between gauge, absolute, and differential pressure sensors?
Gauge sensors measure relative to atmospheric pressure (zero output at 1 atm); used for tire pressure, tank level, HVAC. Absolute sensors measure relative to vacuum (zero output at 0 Pa); used for altimeters, barometers, MAP sensors. Differential sensors measure pressure difference between two ports (positive for P1>P2, negative for P1
What is thermal zero offset and how is it compensated?
Thermal zero offset is shift in bridge output at zero pressure caused by temperature-induced resistance imbalance. Specified as +/-0.02% FS/C typical per IEC 61298. Over 100C range, this is 2% FS error. Compensation methods: (1) analog using thermistor in bridge network, (2) digital using on-chip temperature sensor plus polynomial correction (integrated sensors like Honeywell HSC, Bosch BMP390). Digital sensors achieve +/-0.5% TEB over full temperature range versus +/-2% for uncompensated analog sensors.
How do I select bridge excitation voltage?
Higher excitation gives larger output (better SNR) but increases self-heating. Self-heating error = (Vex^2/Rbridge) * thermal_resistance. For 350 Ohm bridge at 10V: P = 100/350 = 286 mW. With 0.1 C/mW thermal resistance, self-heating is 28.6 C, causing 0.6% FS error at 0.02%/C drift. Standard practice: use 5V for silicon bridges rated 5-10V maximum. For battery-powered applications, use minimum recommended voltage (often 3V) to reduce power to <26 mW per Honeywell HSC low-power modes.