Skip to content

PT100/PT1000 Resistance vs Temperature

Calculate PT100 or PT1000 RTD resistance at any temperature using the Callendar-Van Dusen equation. Get resistance and sensitivity per IEC 60751.

Callendar–Van Dusen equation · IEC 60751 coefficientsAPI slug pt100-resistance

Sign in to saveSave

Sensor and circuit

Example
Sensor

1 Ω or more

Circuit
The divider's fixed resistor: from the reference to the ADC node when the RTD is on the low side, from the ADC node to ground when it is on the high side. A value near the RTD's mid-range resistance gives the most even resolution.

1 Ω or more

The voltage across the divider, which is also the ADC full scale (ratiometric)

100 mV to 50 V

ADC resolution in bits
bit

1 to 24 bit

3.3 V220 ΩADCPt100

Convert

°C

−200 °C to 850 °C

Resistance
110Ω
Divider Output
1.098V
ADC Code
1 363
0x553
Sensitivity
0.3879Ω/°C
Resolution
0.311°C/LSB

Transfer curves

Finest resolution 0.146 °C/LSB at −200 °C
50 Ω100 Ω150 Ω200 Ω250 Ω300 Ω−2000200400600 °CResistance
5001 0001 5002 000−2000200400600 °CADC Code

Lookup table

Every 25 °C from −200 °C to 600 °C
CSV (sign in)
TemperatureResistanceDivider OutputADC Code
−200°C18.5Ω256.2mV318
−175°C29.2Ω386.9mV480
−150°C39.7Ω504.7mV626
−125°C50.1Ω611.7mV759
−100°C60.3Ω709.5mV880
−75°C70.3Ω799.4mV992
−50°C80.3Ω882.5mV1 095
−25°C90.2Ω959.5mV1 190
0°C100Ω1.031V1 280
25°C110Ω1.098V1 363
50°C119Ω1.161V1 440
75°C129Ω1.22V1 513
100°C139Ω1.275V1 582
125°C148Ω1.327V1 646
150°C157Ω1.376V1 707
175°C167Ω1.422V1 765
200°C176Ω1.466V1 819
225°C185Ω1.507V1 871
250°C194Ω1.547V1 919
275°C203Ω1.584V1 966
300°C212Ω1.62V2 010
325°C221Ω1.653V2 052
350°C230Ω1.686V2 092
375°C238Ω1.716V2 130
400°C247Ω1.746V2 166
425°C256Ω1.774V2 201
450°C264Ω1.801V2 234
475°C273Ω1.826V2 266
500°C281Ω1.851V2 297
525°C289Ω1.874V2 326
550°C297Ω1.897V2 354
575°C306Ω1.919V2 381
600°C314Ω1.94V2 407

Formula

IEC 60751 / ITS-90
R(T)=R0(1+AT+BT2)forT≥0°CR(T) = R₀(1 + AT + BT²) \mathrm{for} T ≥ 0°C
SymbolQuantityYour value
R₀Resistance at 0°C (Ω)100 Ω
A3.9083 × 10⁻³ (/°C)—
B−5.775 × 10⁻⁷ (/°C²)—
C−4.183 × 10⁻¹² (below 0 °C only) (/°C⁴)—

Assumptions

6 items
  1. 01Resistance follows the Callendar–Van Dusen equation with IEC 60751's A = 3.9083 × 10⁻³ /°C and B = −5.775 × 10⁻⁷ /°C²: R = R₀(1 + AT + BT²) from 0 °C up, with the term C(T − 100)T³, C = −4.183 × 10⁻¹² /°C⁴, added below 0 °C.
  2. 02R₀ only scales the curve, so a Pt500 or Pt1000 has the shape of a Pt100, and the temperature input spans −200 to 850 °C.
  3. 03Sensitivity is the slope of the curve, R₀(A + 2BT), with the C term's derivative R₀·C·(4T³ − 300T²) added below 0 °C; the resolution uses the same slope.
  4. 04The RTD and the fixed resistor divide the ADC reference, which is also the ADC full scale, so the reading is ratiometric: V_out = V_ref × R/(R + R_fixed) with the RTD on the low side, V_ref × R_fixed/(R + R_fixed) on the high side.
  5. 05The ADC is an ideal unipolar N-bit converter with no offset, gain error, INL, DNL or noise: code = floor(V_out/V_ref × 2^N), held between 0 and 2^N − 1; a code typed in is read at the middle of its step, (code + 0.5)/2^N × V_ref, and solved back to temperature on the same curve.
  6. 06Resolution is (R + R_fixed)²/(R_fixed × 2^N × dR/dT) in °C per step; self-heating from the divider current, the resistance of the leads and loading by the ADC input are not modelled.

Worked example

Pt1000 at 100 °C with a 1 kΩ resistor

A Pt1000 sits on the low side of a divider with a 1 kΩ reference resistor across 3.3 V, and a 12-bit ADC reads the node against the same 3.3 V. The process is at 100 °C.

StepExpressionResult
Resistance at 100 °C1 kΩ × (1 + 0.0039083 × 100 − 5.775×10⁻⁷ × 100²)1 385.055 Ω
Slope of the curve there1 kΩ × (0.0039083 − 2 × 5.775×10⁻⁷ × 100)3.793 Ω/°C
Divider voltage3.3 V × 1 385.055 / (1 385.055 + 1 000)1.916 V
ADC codefloor(1.916 / 3.3 × 2^12)2 378
Resolution at 100 °C(1 385.055 + 1 000)² / (1 000 × 2^12 × 3.793)0.3662 °C/LSB

How it works

This calculator computes PT100/PT1000 resistance from temperature using the IEC 60751:2022 Callendar-Van Dusen equation, essential for calibration technicians, test engineers, and instrumentation designers who need to verify RTD sensor accuracy or design signal conditioning circuits. The resistance-temperature relationship is R(T) = R0*(1 + A*T + B*T^2) for T >= 0 C and R(T) = R0*(1 + A*T + B*T^2 + C*(T-100)*T^3) for T < 0 C. IEC 60751 specifies exact coefficients: A = 3.9083e-3 /C, B = -5.775e-7 /C^2, C = -4.183e-12 /C^4. The sensitivity dR/dT = R0*(A + 2*B*T), plus R0*C*(4*T^3 - 300*T^2) below 0 C, equals 0.405 Ohm/C at -100 C, 0.391 Ohm/C at 0 C, and 0.379 Ohm/C at +100 C for PT100. IEC 60751 accuracy classes define tolerance bands: Class AA is +/-(0.1 + 0.0017*|T|) C, Class A is +/-(0.15 + 0.002*|T|) C, Class B is +/-(0.3 + 0.005*|T|) C. At 0 C, Class AA allows +/-0.04 Ohm deviation from 100.00 Ohm for PT100 sensors.

Practical tips

  • Use 4-wire (Kelvin) connection to eliminate lead resistance errors; even 0.1 Ohm lead resistance introduces 0.26 C error in a PT100 system per ASTM E1137 measurement guidelines
  • Choose PT1000 over PT100 when lead resistance is unavoidable (long cable runs) since lead resistance error is proportionally 10x smaller; a 10 Ohm lead causes only 0.26 C error in PT1000 vs 2.6 C in PT100
  • Limit excitation current to 1 mA or less to keep self-heating below 0.05 C in typical industrial installations per IEC 60751 Annex C recommendations

Common mistakes

  • Using only the two-term CVD equation below 0 C omits the cubic term C, causing errors of 0.02 C at -50 C, 0.21 C at -100 C, and 2.3 C at -200 C per IEC 60751 Annex B verification tables
  • Confusing the IEC/DIN alpha = 0.00385055 with the older ASTM/US standard alpha = 0.003916; using the wrong coefficient set causes 0.3 C error at 100 C, growing to 1.2 C at 400 C
  • Ignoring self-heating: a 1 mA excitation through a 100 Ohm PT100 dissipates 0.1 mW, raising sensor temperature by 0.1-0.5 C depending on thermal coupling to the measured medium

Frequently asked questions

3 questions
What is the difference between PT100 and PT1000?

Both use identical platinum resistance-temperature curves per IEC 60751. PT100 has R0 = 100 Ohm and sensitivity 0.385 Ohm/C; PT1000 has R0 = 1000 Ohm and sensitivity 3.85 Ohm/C. PT1000 provides 10x better resolution for microcontroller ADC interfacing (0.1 C resolution with 12-bit ADC vs 1 C for PT100 on 3.3 V reference) and 10x lower lead resistance sensitivity. Cost difference is minimal; PT1000 is recommended for new designs per TE Connectivity and Honeywell sensor application notes.

What accuracy class should I specify?

IEC 60751 defines Class AA (+/-0.1 C at 0 C), Class A (+/-0.15 C), Class B (+/-0.3 C), and Class C (+/-0.6 C). Class B is sufficient for industrial HVAC, process control, and general automation (+/-0.8 C at 100 C). Class A is used for pharmaceutical manufacturing (FDA 21 CFR Part 211) and food processing (HACCP compliance). Class AA is required for calibration references, metrology labs, and scientific instrumentation per NIST traceability requirements.

Can I use the CVD equation for all RTD materials?

No. The CVD equation with IEC 60751 coefficients applies only to pure platinum RTDs with alpha = 0.00385055. Nickel RTDs use a different polynomial (DIN 43760), with sensitivity 0.617 Ohm/C at 0 C but limited range (-60 to +180 C). Copper RTDs are linear (alpha = 0.00427) but limited to -50 to +150 C. For cryogenic applications below -200 C, use rhodium-iron or germanium RTDs with NIST-calibrated polynomial coefficients.