Thermocouple Voltage & Temperature
Calculate thermocouple EMF voltage from temperature with cold junction compensation. Determine Seebeck output for Type K, J, T, and E sensors.
Constant Seebeck coefficient approximation (K, J, T, E)API slug thermocouple-voltage
Sensor and circuit
Convert
−270 °C to 1 370 °C
- Output Voltage
- 7.175mV
- Amplifier Output
- +430.5mV
- ADC Code
- 534
- 0x216
- Cold Junction Correction
- +1.025mV
- Seebeck Coefficient
- 41µV/°C
- Temperature Difference
- +175°C
- Resolution
- 0.3275°C/LSB
Transfer curves
Resolution 0.3275 °C/LSB across the rangeLookup table
Every 50 °C from 0 °C to 1 250 °C| Hot Junction Temperature | Output Voltage | Amplifier Output | ADC Code |
|---|---|---|---|
| 0°C | −1.025mV | −61.5mV | 0 |
| 50°C | 1.025mV | +61.5mV | 76 |
| 100°C | 3.075mV | +184.5mV | 229 |
| 150°C | 5.125mV | +307.5mV | 381 |
| 200°C | 7.175mV | +430.5mV | 534 |
| 250°C | 9.225mV | +553.5mV | 687 |
| 300°C | 11.28mV | +676.5mV | 839 |
| 350°C | 13.32mV | +799.5mV | 992 |
| 400°C | 15.38mV | +922.5mV | 1 145 |
| 450°C | 17.43mV | +1.045V | 1 297 |
| 500°C | 19.47mV | +1.168V | 1 450 |
| 550°C | 21.52mV | +1.291V | 1 603 |
| 600°C | 23.57mV | +1.414V | 1 755 |
| 650°C | 25.62mV | +1.537V | 1 908 |
| 700°C | 27.67mV | +1.66V | 2 061 |
| 750°C | 29.72mV | +1.783V | 2 213 |
| 800°C | 31.77mV | +1.906V | 2 366 |
| 850°C | 33.83mV | +2.03V | 2 519 |
| 900°C | 35.87mV | +2.152V | 2 671 |
| 950°C | 37.92mV | +2.276V | 2 824 |
| 1 000°C | 39.97mV | +2.398V | 2 977 |
| 1 050°C | 42.02mV | +2.522V | 3 129 |
| 1 100°C | 44.07mV | +2.644V | 3 282 |
| 1 150°C | 46.13mV | +2.768V | 3 435 |
| 1 200°C | 48.17mV | +2.89V | 3 587 |
| 1 250°C | 50.22mV | +3.014V | 3 740 |
Formula
NIST Monograph 175
| Symbol | Quantity | Your value |
|---|---|---|
| S | Seebeck coefficient (K: 41 μV/°C) (μV/°C) | 41 µV/°C |
| T | Temperature (°C) | — |
Assumptions
6 items
- 01Each type has one constant Seebeck coefficient, K 41, J 51, T 43 and E 68 µV/°C, so E = S × (T_hot − T_cold); the NIST ITS-90 reference polynomials are not used, so the result departs from the reference tables wherever the real coefficient differs from the constant.
- 02The cold-junction correction is S × T_cold, the voltage to add to the measured EMF to refer it to a cold junction at 0 °C, with the same constant coefficient.
- 03The amplifier is an ideal gain G with no offset, noise, input bias current or bandwidth limit; the resistance of the leads, connector junctions and the cold-junction sensor's own error are not modelled.
- 04The ADC is an ideal unipolar N-bit converter with full scale V_ref and no INL, DNL or noise: code = floor(V_amp/V_ref × 2^N), held between 0 and 2^N − 1, so a hot junction colder than the cold junction reads 0 and an amplifier output at or above V_ref reads 2^N − 1.
- 05A code typed in is read at the middle of its step, (code + 0.5)/2^N × V_ref.
- 06Resolution is V_ref/(2^N × G × S) in °C per step, the same at every temperature because the model is linear.
Worked example
Type K at 500 °C through a gain-100 amplifier
A type K thermocouple in a 500 °C kiln has its cold junction at a 20 °C terminal block, and a gain-100 amplifier drives a 3.3 V, 12-bit ADC.
| Step | Expression | Result |
|---|---|---|
| Temperature difference across the wires | 500 °C − 20 °C | +480 °C |
| Thermocouple EMF | 41 µV/°C × 480 °C | 19.68 mV |
| Cold-junction correction | 41 µV/°C × 20 °C | +820 µV |
| Amplifier output | 100 × 19.68 mV | +1.968 V |
| ADC code | floor(1.968 / 3.3 × 2^12) | 2 442 |
| Resolution | 3.3 V / (2^12 × 100 × 41 µV/°C) | 0.1965 °C/LSB |
How it works
This calculator computes thermocouple EMF from temperature using the Seebeck effect, essential for process engineers, instrumentation technicians, and control system designers measuring temperatures from -270 to +2300 C. Thermocouples generate a voltage proportional to the temperature difference between the hot (measurement) and cold (reference) junctions: E = S * (T_hot - T_cold), where S is the Seebeck coefficient in uV/C. NIST ITS-90 thermocouple tables (Monograph 175) define standard coefficients: Type K (Chromel-Alumel) = 41 uV/C, Type J (Iron-Constantan) = 51 uV/C, Type T (Copper-Constantan) = 43 uV/C, Type E (Chromel-Constantan) = 68 uV/C (highest sensitivity). Cold junction compensation (CJC) is mandatory since the reference junction is at instrument temperature, not 0 C. The linear Seebeck approximation provides +/-2-3% accuracy over 100 C spans; for precision applications, NIST polynomial tables achieve +/-0.02 C accuracy per IEC 60584-1:2013.
Practical tips
- Use the same alloy extension wire as the thermocouple (Type K extension with Type K sensor) to avoid creating additional Seebeck junctions at connectors per ASTM E230 requirements
- For temperatures above 1000 C, Type K accuracy degrades due to chromium oxidation; switch to Type R or S (platinum-rhodium) for +/-0.25% accuracy up to 1600 C per IEC 60584-2
- Dedicated thermocouple amplifier ICs (AD8495, MAX31855) include integrated CJC and provide direct digital output, simplifying signal conditioning to a single component
Common mistakes
- Ignoring cold junction compensation: if the terminal is at 30 C instead of 0 C, the error is 30*41 = 1230 uV = 30 C temperature error for Type K; modern instruments include automatic CJC but older meters may not
- Using wrong thermocouple type calibration: Type K and Type J cables look identical; applying J calibration to K wire causes errors up to 50 C at 800 C per IEC 60584-1 deviation tables
- Routing thermocouple extension wire near power cables: the millivolt signals couple inductively; IEEE 518 requires minimum 50 mm separation or use of twisted shielded thermocouple extension wire
Frequently asked questions
3 questions
Why is cold junction compensation necessary?
Thermocouples measure temperature difference between hot and cold junctions, not absolute temperature. If the cold junction (instrument terminal) is not at 0 C, the reading includes an offset. CJC measures terminal temperature with a separate sensor (thermistor or RTD) and adds the equivalent voltage correction. Without CJC, a 25 C terminal temperature causes 25 C measurement error. Modern transmitters include automatic CJC; verify it is enabled and calibrated per ISA-MC96.1 requirements.
What thermocouple type is best for general-purpose use?
Type K (Chromel-Alumel) is the industry standard, covering -200 to +1372 C with 41 uV/C sensitivity and +/-2.2 C or +/-0.75% accuracy (whichever is greater) per IEC 60584-1 Class 2. It represents 80% of industrial thermocouple installations. Type T is preferred for low temperatures (-200 to +350 C) due to better accuracy (+/-1 C) and oxidation resistance in humid environments. Type J offers higher sensitivity (51 uV/C) but is limited to +760 C maximum.
How accurate is the linear Seebeck approximation?
The linear approximation is +/-2-3% over a 100 C span around the calibration point, adequate for most industrial monitoring. For precision measurement or wide temperature ranges, NIST Monograph 175 provides polynomial coefficients achieving +/-0.02 C accuracy. The polynomial has 9-14 terms depending on thermocouple type and temperature range. Microcontroller implementations typically use lookup tables with linear interpolation for +/-0.1 C accuracy with minimal computation.