Voltage Unit Converter
Convert voltage between microvolts, millivolts, volts, kilovolts, and megavolts.
SI prefixes · 1 V = 10⁶ µV = 10³ mV = 10⁻³ kV = 10⁻⁶ MVAPI slug voltage-units
Linked values
Type in any field — the others follow.
Reference values
Select a row to load it| Microvolts | Millivolts | Voltage | Kilovolts | Megavolts |
|---|---|---|---|---|
| µV | 0.001mV | 1.000e−6V | 1.000e−9kV | 1.000e−12MV |
| µV | 0.010mV | 1.000e−5V | 1.000e−8kV | 1.000e−11MV |
| µV | 0.100mV | 1.000e−4V | 1.000e−7kV | 1.000e−10MV |
| µV | 1.000mV | 0.001V | 1.000e−6kV | 1.000e−9MV |
| µV | 10.000mV | 0.010V | 1.000e−5kV | 1.000e−8MV |
| µV | 100.000mV | 0.100V | 1.000e−4kV | 1.000e−7MV |
| µV | 1000.000mV | 1.000V | 0.001kV | 1.000e−6MV |
| µV | 1.000e6mV | 1000.000V | 1.000kV | 0.001MV |
Formula
| Symbol | Quantity | Your value |
|---|
Assumptions
4 items
- 01The input is in volts and every other unit is a decimal SI multiple: 1 V = 10⁶ µV = 10³ mV, 1 kV = 10³ V and 1 MV = 10⁶ V, exactly.
- 02The value is one voltage taken as given; DC, RMS, peak or peak-to-peak is kept, and no conversion between them is made.
- 03Logarithmic units such as dBV and dBµV are not converted here.
- 04Voltage is entered from 0 V upward with no upper limit; the field refuses a negative value, so enter the magnitude and keep the polarity separately.
Worked example
Type K thermocouple at 100 °C
A type K thermocouple produces 4.096 mV at 100 °C with its reference junction at 0 °C (IEC 60584-1 table). The ADC driver is configured in volts, and the amplifier noise budget is in microvolts.
| Step | Expression | Result |
|---|---|---|
| Volts | 4.096 mV ÷ 10³ | 0.004096 V |
| Microvolts | 0.004096 V × 10⁶ | 4 096 µV |
| Average sensitivity from 0 °C | 4 096 µV ÷ (100 °C − 0 °C) | 40.96 µV/°C |
How it works
This calculator converts between volts, millivolts, microvolts, and kilovolts for electronics engineers, instrumentation designers, and power systems professionals. Per SI Brochure (BIPM), the volt is defined as W/A = J/C = kg·m^2/(A·s^3), traceable to Josephson voltage standard with 10^-10 relative uncertainty. Electronics spans 12 orders of magnitude: nanovolts for SQUID magnetometers, microvolts for thermocouples (40 uV/C for Type K per IEC 60584), millivolts for Wheatstone bridges (2 mV/V typical), volts for logic levels (3.3 V LVCMOS per JEDEC), and kilovolts for ESD testing (8 kV per IEC 61000-4-2). Thermal noise in a 1 MHz bandwidth at 50 ohm: 0.91 uV RMS per Johnson-Nyquist equation.
Practical tips
- Thermal noise voltage per Johnson-Nyquist: V_n = sqrt(4kTRB) where k = 1.380649 × 10^-23 J/K. At 290 K, 50 ohm, 1 MHz: V_n = 0.91 uV RMS. This sets fundamental SNR limit for sensitive measurements
- Logic voltage levels per JEDEC: LVTTL Vih > 2.0 V, Vol < 0.4 V; LVCMOS 3.3 V Vih > 2.0 V, Vol < 0.4 V; LVCMOS 1.8 V Vih > 1.17 V, Vol < 0.45 V. Verify both high and low thresholds for reliable interfacing
- ESD test levels per IEC 61000-4-2: contact discharge 2-8 kV, air discharge 2-15 kV. A 2 kV ESD pulse contains ~0.5 mJ but delivers 7.5 A peak current - enough to damage 3.3 V CMOS gates
Common mistakes
- Confusing mV (10^-3 V) with uV (10^-6 V) - they differ by 1000x. Thermocouple output is mV-range; amplifier input noise is uV-range. A 10 uV noise on 4 mV signal = 0.25% error
- Ignoring voltage drop in high-current paths - per IPC-2221, a 10 A current through 10 mohm trace resistance causes 100 mV drop, significant for 3.3 V logic rails (3% drop)
- Using oscilloscope with wrong vertical scale - mixing up mV/div and V/div leads to 1000x amplitude error. A 5 mV signal on 5 V/div scale appears as flat line
Frequently asked questions
4 questions
What is a typical noise floor in microvolts?
Thermal (Johnson) noise at 290 K: V_n = 0.13 × sqrt(R × BW_MHz) uV. For 50 ohm, 1 MHz: 0.91 uV RMS. Op-amp input noise: 1-20 nV/sqrt(Hz) typical, so 3 uV RMS in 100 kHz bandwidth. Low-noise instrumentation amplifiers (AD8429): 1 nV/sqrt(Hz) = 0.3 uV in 100 kHz BW.
When do I use millivolts vs volts on a multimeter?
Use mV range for: thermocouple output (0-50 mV), shunt resistor voltage (10-100 mV at rated current), bridge sensor output (0-30 mV), small voltage drops. Use V range for: power supply rails, logic signals, battery voltages. DMM resolution: 4.5-digit on mV range = 1 uV, on V range = 1 mV.
What is the difference between RMS and peak voltage?
For sine wave: V_peak = V_RMS × sqrt(2) = 1.414 × V_RMS per IEEE definition. Mains 230 V RMS = 325 V peak. For power calculations: P = V_RMS^2/R (true for any waveform). Multimeters display true RMS for AC; cheap meters assume sine wave and can have 40% error on non-sinusoidal waveforms.
Why are ESD test voltages specified in kV?
Human body model (HBM) per JEDEC JESD22-A114: 2-8 kV discharge through 1.5 kohm, 100 pF. Despite high voltage, energy is low (~0.2-3.2 uJ). Damage occurs from current density: 8 kV HBM produces 5.3 A peak current. Gate oxide breakdown in 3.3 V CMOS occurs at ~7 V, so 8 kV easily punches through multiple oxide layers.