NTC Thermistor Temperature Calculator
Calculate temperature from NTC thermistor resistance using the Steinhart-Hart beta equation. Get Kelvin and Celsius outputs for sensor circuit design.
β-parameter equation · 1/T = 1/T₀ + ln(R/R₀)/βAPI slug ntc-thermistor
Sensor and circuit
Convert
1 Ω to 100 MΩ
- Temperature
- 25°C
- Divider Output
- 1.65V
- ADC Code
- 2 048
- 0x800
- Temperature (Kelvin)
- 298.1K
- Resolution
- 0.02198°C/LSB
Transfer curves
Finest resolution 0.0215 °C/LSB at 19.1 °CLookup table
Every 5 °C from −40 °C to 125 °C| Temperature | Measured Resistance | Divider Output | ADC Code |
|---|---|---|---|
| −40°C | 401.9kΩ | 3.22V | 3 996 |
| −35°C | 281.6kΩ | 3.187V | 3 955 |
| −30°C | 200.2kΩ | 3.143V | 3 901 |
| −25°C | 144.3kΩ | 3.086V | 3 830 |
| −20°C | 105.4kΩ | 3.014V | 3 741 |
| −15°C | 77.9kΩ | 2.925V | 3 630 |
| −10°C | 58.25kΩ | 2.816V | 3 495 |
| −5°C | 44.03kΩ | 2.689V | 3 337 |
| 0°C | 33.62kΩ | 2.543V | 3 156 |
| 5°C | 25.92kΩ | 2.381V | 2 955 |
| 10°C | 20.17kΩ | 2.206V | 2 738 |
| 15°C | 15.84kΩ | 2.023V | 2 510 |
| 20°C | 12.54kΩ | 1.836V | 2 278 |
| 25°C | 10kΩ | 1.65V | 2 048 |
| 30°C | 8.037kΩ | 1.47V | 1 825 |
| 35°C | 6.506kΩ | 1.301V | 1 614 |
| 40°C | 5.301kΩ | 1.143V | 1 419 |
| 45°C | 4.348kΩ | 1V | 1 241 |
| 50°C | 3.588kΩ | 871.4mV | 1 081 |
| 55°C | 2.978kΩ | 757.3mV | 939 |
| 60°C | 2.486kΩ | 657.1mV | 815 |
| 65°C | 2.086kΩ | 569.7mV | 707 |
| 70°C | 1.76kΩ | 493.8mV | 612 |
| 75°C | 1.492kΩ | 428.4mV | 531 |
| 80°C | 1.27kΩ | 372mV | 461 |
| 85°C | 1.087kΩ | 323.5mV | 401 |
| 90°C | 933.6Ω | 281.8mV | 349 |
| 95°C | 805.4Ω | 246mV | 305 |
| 100°C | 697.5Ω | 215.2mV | 267 |
| 105°C | 606.4Ω | 188.7mV | 234 |
| 110°C | 529.1Ω | 165.8mV | 205 |
| 115°C | 463.3Ω | 146.1mV | 181 |
| 120°C | 407.1Ω | 129.1mV | 160 |
| 125°C | 358.8Ω | 114.3mV | 141 |
Formula
| Symbol | Quantity | Your value |
|---|---|---|
| T | Temperature (K) | 25 °C |
| T₀ | Reference temperature (K) | 25 °C |
| β | Beta coefficient (K) | 3 950 K |
| R | Measured resistance (Ω) | 10 kΩ |
| R₀ | Reference resistance at T₀ (Ω) | 10 kΩ |
Assumptions
6 items
- 01The thermistor follows the β equation, 1/T = 1/T₀ + ln(R/R₀)/β, with one β over the whole range and 273.15 between kelvin and °C; the Steinhart–Hart third-order term is not used, so far from T₀ the result is only as good as one β fits the part.
- 02The NTC and the fixed resistor divide the ADC reference, which is also the ADC full scale, so the reading is ratiometric: V_out = V_ref × R/(R + R_fixed) with the NTC on the low side, V_ref × R_fixed/(R + R_fixed) on the high side.
- 03The ADC is an ideal unipolar N-bit converter with no offset, gain error, INL, DNL or noise, and its input draws no current from the divider: code = floor(V_out/V_ref × 2^N), held between 0 and 2^N − 1.
- 04A code typed in is read at the middle of its step, (code + 0.5)/2^N × V_ref, and turned back into resistance and temperature through the same divider and β equation.
- 05Resolution is the local slope T²(R + R_fixed)²/(β × R × R_fixed × 2^N), in kelvin (the same as °C) per step, and is the same with the NTC on either side.
- 06Self-heating from the divider current and the resistance of the leads are not modelled.
Worked example
Code 1 000 from a 10 kΩ NTC
Firmware on a 3.3 V, 12-bit MCU reads code 1 000 from a 10 kΩ, β 3 950 NTC on the low side of a divider, with a 10 kΩ resistor up to the 3.3 V reference. Read at the middle of that step, what temperature is it?
| Step | Expression | Result |
|---|---|---|
| Voltage at the middle of code 1 000 | (1 000 + ½) / 2^12 × 3.3 V | 806.1 mV |
| Thermistor resistance | 10 kΩ × 0.8061 / (3.3 − 0.8061) | 3.232 kΩ |
| Temperature from the β equation | 1 / (1/(25 + 273.15) + ln(3 232 / 10 000) / 3 950) − 273.15 | 52.79 °C |
| Code the ADC reads there | floor(0.8061 / 3.3 × 2^12) | 1 000 |
| Resolution at this point | 325.9² × (3 232 + 10 000)² / (3 950 × 3 232 × 10 000 × 2^12) | 0.03557 °C/LSB |
How it works
This calculator converts NTC thermistor resistance to temperature using the Steinhart-Hart equation, essential for embedded systems engineers, IoT developers, and industrial control designers who need accurate temperature sensing from -40 to +125 C. NTC (Negative Temperature Coefficient) thermistors decrease resistance as temperature rises, following R(T) = R25 * exp(B * (1/T - 1/298.15)), where B is the material constant (typically 3000-5000 K per manufacturer datasheets). According to IEC 60539-1, standard NTC thermistors achieve +/-1% resistance tolerance at 25 C, translating to +/-0.2 C accuracy. The Steinhart-Hart three-coefficient model (a + b*ln(R) + c*ln(R)^3 = 1/T) provides +/-0.02 C accuracy across the full range per NIST calibration guidelines. Industrial-grade NTCs from Vishay, Murata, and TDK specify B-values with +/-1% tolerance, yielding +/-0.5 C measurement uncertainty over -40 to +85 C operating range. Response time (tau63) ranges from 0.5 s for bare-chip sensors to 15 s for encapsulated probes in still air per IEC 60539-2 test methods.
Practical tips
- Use manufacturer lookup tables or Steinhart-Hart coefficients from the datasheet for +/-0.1 C accuracy; the simplified B-equation is only +/-1 C accurate per NIST Technical Note 1297
- Limit excitation current to 100 uA for precision measurement to keep self-heating below 0.01 C per IEC 60539-2 recommendations
- For linearization, add a parallel resistor equal to R25 to achieve +/-3% linearity over +/-25 C span around the center point per Vishay application note AN-NTCS-1
Common mistakes
- Using a generic B-value (3950 K) instead of the datasheet-specific value causes +/-2-5 C errors at temperature extremes; Murata NCP series specifies B25/50 vs B25/85 separately with up to 3% difference
- Forgetting Kelvin conversion: using 25 instead of 298.15 K in the equation produces nonsensical negative temperatures or overestimates by 10-20 C
- Ignoring self-heating: 1 mA through a 10 kOhm NTC at 25 C dissipates 10 mW, raising sensor temperature by 0.1-1.0 C depending on thermal coupling per IEC 60539-1 dissipation constant specification
Frequently asked questions
2 questions
What does NTC mean and how does it differ from PTC?
NTC (Negative Temperature Coefficient) resistance decreases 3-5%/C as temperature rises, following an exponential curve. PTC (Positive Temperature Coefficient) resistance increases with temperature. NTCs are preferred for temperature measurement due to their 10x higher sensitivity (typically -4%/C vs +0.4%/C for platinum RTDs per IEC 60751), while PTCs are used for overcurrent protection where resistance sharply increases above a threshold.
How accurate are NTC thermistors compared to RTDs?
NTCs achieve +/-0.1 to +/-1 C accuracy depending on calibration, comparable to Class B RTDs (+/-0.3 C at 0 C per IEC 60751). Key differences: NTCs have 10x higher sensitivity (better resolution), lower cost (USD 0.10-2 vs USD 5-50 for RTDs), but narrower range (-40 to +125 C vs -200 to +850 C for platinum RTDs). For industrial temperature measurement from -40 to +150 C, interchangeable NTCs meeting IEC 60539-1 provide +/-0.2 C accuracy at a fraction of RTD cost.